Climate Chamber – TE-4001

Temperature range

20°C to 60°C

Temperature controller

Microprocessor-based digital PID controller with RBC (Brazilian Calibration Network) calibration certificate

Sensor

PT-100 probe

Control accuracy

±0,1°C

Uniformity

±1°C

Working ranges

Temperature: 30ºC.........Humidity: 60%
Temperature: 30°C.........Humidity: 75%
Temperature: 40ºC.........Humidity: 75%
Temperature: 60ºC.........Humidity: 90%

Humidity

25% to 95%

Humidity controller

Microprocessor-based digital PID controller with RBC (Brazilian Calibration Network) calibration certificate

Humidity sensor

Vaisala INTERCAP

Control accuracy

±1%

Uniformity

±5%

Boiler

In borosilicate glass with drain, electronic float switch and system for protection in case of water shortage

Compressor

1/3HP hermetic compressor with CFC-free R-134-A gas

Cooling capacity

2,200 BTU/hour at 5°C

Internal airflow

Forced circulation

Capacity

07 shelves 120 mm apart from each other

Safety

Safety thermostat with auto shut down in case of overheating

Inner chamber

In polished 304 stainless steel

Inner door

In tempered glass

Door

In carbon steel with outer lock, safety switch and resistance to avoid condensation

Casters

Swivel casters with locking system

Body

In carbon steel with electrostatic coating and anti-corrosion treatment

Inner dimensions

W=535 x D=475 x H=1,000 mm

Volume

250 Liters

Outer dimensions

W=705 x D=850 x H=1,745 mm

Weight

185 kg

Power

2800 Watts

Voltage

220 Volts

Parts included

- One reverse osmosis device, which is able to retain physical, organic, chemical and microbiological concentrations from water, thus avoiding concentration of mineral salts inside the boiler of the humidity system;
- seven stainless steel perforated trays;
- User manual with warranty deed.

Optional

RS-485 serial communication for temperature and humidity monitoring via software